| Safety Standards Measuring Instrument (AC Withstand Voltage Testers) |
AUTOMATIC INSULATION / WITHSTANDING HiTESTER 3153 |
|
 |
|
 |
 |
|
|
 |
 |
| Programable testing, full remote control, Automatic Insulation Withstanding Tester |
 |
This electrical safety test instrument series is designed for insulation resistance and voltage withstand testing of electrical devices and components according to various safety standards. A multitude of automation and labor-saving features are provided to ensure effective testing for a wide variety of requirements and test conditions. Select the most appropriate model for your applications.
- Insulation and AC/DC continuous voltage withstand testing
- Safety standards compliance testing
- Stores up to 10 sets of test conditions (10 each for insulation and voltage withstand)
- Zero-V Switching (voltage withstand testing)
- Electrical Discharge function (DC voltage withstand, insulation testing)
- Automated testing support
- Interlock Protect function
- Remote controllable
|
 |
| Basic specifications : Withstanding test section |
 |
| Testing voltage |
0.2 kV to 5.00 kV AC, 500VA (maximum 30 minutes) 0.2 kV to 5.00 kV DC, 50VA (continuance) |
| Voltage setting method |
Digital setting |
| Waveform |
Same as the power supply waveform |
| Frequency |
Same as the power supply frequency |
| Measurement range |
Current: 0.01 to 100.0 mA, ±(2 % f.s. +5 dgt.) 10 mA/100 mA AC (Average value rectified, effective value display) |
|
|
| Basic specifications : Insulation test section |
 |
| Testing voltage |
50 V to 1000 V DC |
| Measurement range |
0.1 m-ohm to 9999 m-ohm, 4 ranges |
|
|
| Basic specifications : general |
 |
| Judgment function |
Contents: UPPER-FAIL / PASS / LOWER-FAIL (Digital setting window comparator method) |
| Timer section |
Setting range: 0.3 to 999 seconds |
| Interfaces |
EXT I/O, EXT SW, RS-232C, GP-IB |
| Display |
Fluorescent tube display (digital), Analog meter |
| Monitor function |
Output voltage, detection current |
| Power supply |
100 V to 120 V AC / 200 to240 V AC, (50/60 Hz), 1000 VA max. |
| Dimensions, mass |
320 mm(12.60 in)W × 155 mm(6.10 in)H × 480 mm(18.90 in)D, 18 kg (634.9 oz) |
| Accessories |
H.V. TEST LEAD 9615 (high voltage side and return, 1 each), Power cord (1), spare fuse (1) |
|
|