HIOKI has decided to take steps to revise the accuracy spec. regarding the measurement at the low temperature ( –35.0 to –0.1°C) on the Models INFRARED THERMO HiTESTER 3419 and 3419-20. We apologize for any inconvenience caused by this change, and thank you for your kind understanding and cooperation.


1. Model name:
  • INFRARED THERMO HiTESTER 3419 and 3419-20
2. Details of the revised spec.

Current spec.

Accuracy ±2% rdg. or ±2°C, whichever value is higher

Revised spec.
Accuracy
-35.0 to -0.1°C :  ±10%rdg.±2°C
0.0 to 100.0°C :  ±2°C:
100.1 to 500.0°C :  ±2%rdg.
» The accuracy specification regarding the measurement at 0°C or higher temperature does not change but only the expression has changed.

Thank you for your cooperation

Best regards
Atsushi Mizuno