Sep. 29th, 2009
Dear Valued Customers:
Thank you very much for your continued patronage.
HIOKI has decided to take steps to revise the accuracy spec. regarding the measurement at the low temperature ( –35.0 to –0.1°C) on the Models INFRARED THERMO HiTESTER 3419 and 3419-20. We apologize for any inconvenience caused by this change, and thank you for your kind understanding and cooperation.
| 1. |
Model name: |
- INFRARED THERMO HiTESTER 3419 and 3419-20
|
|
| 2. |
Details of the revised spec. |
Current spec.
| Accuracy |
±2% rdg. or ±2°C, whichever value is higher |
Revised spec.
| Accuracy |
| -35.0 to -0.1°C : |
±10%rdg.±2°C |
| 0.0 to 100.0°C : |
±2°C: |
| 100.1 to 500.0°C : |
±2%rdg. |
|
» The accuracy specification regarding the measurement at 0°C or higher temperature does not change but only the expression has changed.
|
|
Thank you for your cooperation
Best regards
Atsushi Mizuno
Director
Quality Assurance Department
HIOKI E.E. CORPORATION |